Máy hiện sóng Instek GOS-6112 - 100MHz Analog Oscilloscope

GOS-6112, máy hiện sóng tương tự, băng thông 100MHz, 2 kênh, độ nhạy cao của Instek


Vui lòng tham khảo thông số kỹ thuật và bảng giá sản phẩm

        Pricelist



Features:
  • 100MHz Bandwidth
  • Dual Channel
  • Delayed Sweep
  • Cursor Readout with 7 Measurements
  • TV Synchronisation
  • Z-Axis Modulation Input

Specifications
CRT Type: 6-inch  rectangular type with internal graticule; 0%, 10%, 90% and 100% markers; 8 x 10 DIV (1 DIV = 1 cm)
Phosphor: P31
Accelerating Potential: 12kV approx
Z-Axis Input: Coupling: DC
Sensitivity: 5V or more
Maximum input voltage: 30V (DC + AC peak) at 1kHz or less
Bandwidth: DC ~ 5MHz
Vertical System Sensitivity: 2mV ~ 5V/DIV, 11 step in 1-2-5 sequence
Sensitivity Accuracy: £ 3% (5 DIV at the centre of display)
Vernier Vertical Sensitivity: Continuously variable to 1/2.5 or less of panel-indicated value
Bandwidth (-3dB): DC ~ 100MHz (2mV/DIV : DC ~ 20MHz)
Rise Time: 3.5ns (2mV/DIV : 17.5ns)
Signal Delay: Leading edge can be monitored
Maximum Input Voltage: 400V (DC + AC peak) at 1kHz or less
Input Coupling: AC, DC, GND
Input Impedance: 1MW ±2% // approx 25pF
Vertical Modes: CH1, CH2, DUAL (CHOP/ALT), ADD, CH2 INV
Bandwidth Limited: 20MHz
Common Mode Rejection Ratio: 50:1 or better at 50kHz
Dynamic Range: 8 DIV at 60MHz; 5 DIV at 100MHz
Horizontal System Horizontal Modes: MAIN (A), ALT, DELAY (B)
A (Main) Sweep Time: 50ns ~ 0.5s/DIV, continuously variable (UNCAL)
B (Delay) Sweep Time: 50ns ~ 50ms/DIV
Accuracy: ±3% (±5% at x 10 MAG)
Sweep Magnification: x 10 (maximum sweep time 5ns/DIV)
Hold Off Time: Variable
Delay Time: 1µs ~ 5s
Delay Jitter: Better than 1:20000
Alternate Separation: Variable
Trigger Trigger Modes: AUTO, NORM, TV
Trigger Source: CH1, CH2, LINE, EXT
Trigger Coupling: AC, DC, HFR, LFR
Trigger Slope: + or - polarity or TV sync polarity
Trigger Sensitivity:
Mode Frequency INT EXT
AUTO 10Hz ~ 20MHz 0.35 DIV 50mV
20MHz ~ 100MHz 1.5 DIV 150mV
NORM DC ~ 20MHz 0.35 DIV 50mV
20MHz ~ 100MHz 1.5 DIV 150mV
TV Sync signal 1 DIV 200mVpp
TV Sync: TV-V, TV-H
Max. External Input Voltage: 400V (DC + AC peak) at 1kHz
External Input Impedance: 1MW ±5% // approx 25pF
X-Y Operation Mode: X-axis: selectable CH1, EXT; Y-axis: selectable CH1, CH2, CH1 and CH2
Sensitivity Accuracy: 2mV ~ 5V/DIV ±3%; EXT: 0.1V/DIV ±5%
X-Axis Bandwidth: DC ~ 500kHz (-3dB)
Phase Error: 3º or less from DC ~ 50kHz
Output Signal Trigger Signal Output: Voltage: approx 25mV/DIV into 50W; Frequency response: DC ~ 10MHz
Calibrator Output: 1kHz square wave, 2Vpp ±2%
Cursor Readout Function Cursor Measurement Function: DV, DV%, DVdB, DT, 1/DT, DT%, Dq
Cursor Resolution: 1/100 DIV
Effective Cursor Range: Vertical: ±3 DIV; Horizontal: ±4 DIV
Panel Setting Display: Vertical: V/DIV (CH1, CH2), UNCAL, ALT/CHOP/ADD, INV, probe factor, AC/DC/GND
Horizontal: S/DIV (MTB, DTB), UNCAL, x 10 MAG, delay time, HO
Trigger: source, coupling, slope, level, TV-V, TV-H
Others: X-Y, lock
Special Functions Panel Setup Lock: Provided
Other Details Power Source: AC 100V/120V/230V ±10%, 50/60Hz
Accessories: Power cord x 1, Instruction manual x 1, Probe (10:1/1:1) x 2
Dimensions: 310mm (w) x 150mm (h) x 455mm (d)
Weight: 9kg approx